Step-by-step triage: element dropout patterns, connector issues — and when a $150 cable fixes a $2,000 problem.
A dark band on the image does not automatically mean a dead probe. In our repair experience, a meaningful share of "broken probes" are actually cable or connector faults — the cheap part of the assembly. Ten minutes of triage can save a two-thousand-dollar decision.
With the image live, gently flex the cable at three points: at the connector, mid-cable, and at the strain relief where the cable enters the probe body. If the artifact appears or moves as you flex, the fault is in the cable — not the array.
Test the probe on a second machine if possible, or test a known-good probe on the suspect machine. Half of "probe failures" we hear about are actually system-side: a damaged port, a software profile, or a bad peripheral.
| Fault | Usually means | Typical fix |
|---|---|---|
| Artifact moves when cable flexes | Cable conductors | Cable / strain-relief replacement — a fraction of probe cost |
| Intermittent contact at connector | Connector pins | Connector replacement |
| Stable vertical dropout band | Array elements | Probe-level repair or graded replacement probe |
| TEE sheath damage | Infection control | Sheath replacement or graded replacement probe |
We stock probes, cables and connectors, and every used probe we sell is image-tested with the dropout pattern photographed. Not sure which side your fault is on? Send us the probe model, the system model and a photo of the artifact — we will tell you honestly which part you need, even if it is the cheap one.